The following table shows the results of 11 side-by-side blood tests performed using the LifeScan FastTake, the Glucometer Dex, and Accu-Chek Complete meters. The results are in mg/dl.
| Sample | FastTake | Glucometer Dex | Accu-Chek Complete | Variance FT/Dex | Variance FT/Complete |
| 1 | 93 | 92 | 78 | 1% | 19% |
| 2 | 83 | 88 | 75 | -6% | 11% |
| 3 | 62 | 52 | 61 | 19% | 2% |
| 4 | 87 | 90 | 68 | -3% | 28% |
| 5 | 70 | 75 | 71 | -7% | -1% |
| 6 | 237 | 221 | 235 | 7% | 1% |
| 7 | 68 | 69 | 63 | -1% | 8% |
| 8 | 162 | 161 | 151 | 1% | 7% |
| 9 | 179 | 175 | 171 | 2% | 5% |
| 10 | 372 | 309 | 322 | 20% | 16% |
| 11 | 222 | 233 | 236 | -5% | -6% |
| Average | 149 | 142 | 139 | 3% | 8% |
Comparing the results of the FastTake and the Glucometer Dex yields a correlation coefficient of 0.987. Comparing the results of the FastTake and the Accu-Chek Complete yields a correlation coefficient of 0.988.
Return to the review of the: